CVE-2017-12712
| Summary |
The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communications. CVSS v3 base score: 7.5, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
|
| Publication Date |
April 25, 2018, 10:29 p.m. |
| Registration Date |
Jan. 26, 2021, 1:14 p.m. |
| Last Update |
Nov. 21, 2024, 12:10 p.m. |
|
CVSS3.0 : HIGH
|
| スコア |
8.8
|
| Vector |
CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
| 攻撃元区分(AV) |
隣接 |
| 攻撃条件の複雑さ(AC) |
低 |
| 攻撃に必要な特権レベル(PR) |
不要 |
| 利用者の関与(UI) |
不要 |
| 影響の想定範囲(S) |
変更なし |
| 機密性への影響(C) |
高 |
| 完全性への影響(I) |
高 |
| 可用性への影響(A) |
高 |
|
CVSS2.0 : HIGH
|
| Score |
8.3
|
| Vector |
AV:A/AC:L/Au:N/C:C/I:C/A:C |
| 攻撃元区分(AV) |
隣接 |
| 攻撃条件の複雑さ(AC) |
低 |
| 攻撃前の認証要否(Au) |
不要 |
| 機密性への影響(C) |
高 |
| 完全性への影響(I) |
高 |
| 可用性への影響(A) |
高 |
| Get all privileges. |
いいえ
|
| Get user privileges |
いいえ
|
| Get other privileges |
いいえ
|
| User operation required |
いいえ
|
Affected software configurations
| Configuration1 |
or higher |
or less |
more than |
less than |
| cpe:2.3:o:abbott:accent_firmware:*:*:*:*:*:*:*:* |
|
|
|
f0b.0e.7e |
| execution environment |
| 1 |
cpe:2.3:h:abbott:accent:-:*:*:*:*:*:*:* |
| Configuration2 |
or higher |
or less |
more than |
less than |
| cpe:2.3:o:abbott:anthem_firmware:*:*:*:*:*:*:*:* |
|
|
|
f0b.0e.7e |
| execution environment |
| 1 |
cpe:2.3:h:abbott:anthem:-:*:*:*:*:*:*:* |
| Configuration3 |
or higher |
or less |
more than |
less than |
| cpe:2.3:o:abbott:accent_mri_firmware:*:*:*:*:*:*:*:* |
|
|
|
f10.08.6c |
| execution environment |
| 1 |
cpe:2.3:h:abbott:accent_mri:-:*:*:*:*:*:*:* |
| Configuration4 |
or higher |
or less |
more than |
less than |
| cpe:2.3:o:abbott:accent_st_firmware:*:*:*:*:*:*:*:* |
|
|
|
f10.08.6c |
| execution environment |
| 1 |
cpe:2.3:h:abbott:accent_st:-:*:*:*:*:*:*:* |
| Configuration5 |
or higher |
or less |
more than |
less than |
| cpe:2.3:o:abbott:assurity_firmware:*:*:*:*:*:*:*:* |
|
|
|
f14.07.80 |
| execution environment |
| 1 |
cpe:2.3:h:abbott:assurity:-:*:*:*:*:*:*:* |
| Configuration6 |
or higher |
or less |
more than |
less than |
| cpe:2.3:o:abbott:allure_firmware:*:*:*:*:*:*:*:* |
|
|
|
f14.07.80 |
| execution environment |
| 1 |
cpe:2.3:h:abbott:allure:-:*:*:*:*:*:*:* |
| Configuration7 |
or higher |
or less |
more than |
less than |
| cpe:2.3:o:abbott:assurity_mri_firmware:*:*:*:*:*:*:*:* |
|
|
|
f17.01.49 |
| execution environment |
| 1 |
cpe:2.3:h:abbott:assurity_mri:-:*:*:*:*:*:*:* |
Related information, measures and tools
Common Vulnerabilities List
JVN Vulnerability Information
Abbott Laboratories pacemakers における暗号に関する脆弱性
| Title |
Abbott Laboratories pacemakers における暗号に関する脆弱性
|
| Summary |
Abbott Laboratories pacemakers には、暗号に関する脆弱性が存在します。
|
| Possible impacts |
情報を取得される、情報を改ざんされる、およびサービス運用妨害 (DoS) 状態にされる可能性があります。 |
| Solution |
ベンダ情報および参考情報を参照して適切な対策を実施してください。 |
| Publication Date |
Aug. 29, 2017, midnight |
| Registration Date |
June 22, 2018, 1:50 p.m. |
| Last Update |
July 4, 2018, 3:54 p.m. |
Affected System
| アボット |
|
Accent
|
|
Accent MRI
|
|
Accent ST
|
|
Allure
|
|
Anthem
|
|
Assurity
|
|
Assurity MRI
|
CVE (情報セキュリティ 共通脆弱性識別子)
CWE (共通脆弱性タイプ一覧)
ベンダー情報
その他
Change Log
| No |
Changed Details |
Date of change |
| 1 |
[2018年06月22日] 掲載 |
June 22, 2018, 1:50 p.m. |
| 2 |
[2018年07月04日] 参考情報:ICS-CERT ADVISORY (ICSMA-18-107-01) を追加 |
July 4, 2018, 3:50 p.m. |