| Summary | A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation. |
|---|---|
| Publication Date | Feb. 15, 2018, 7:29 p.m. |
| Registration Date | Jan. 26, 2021, 1:14 p.m. |
| Last Update | Nov. 21, 2024, 12:10 p.m. |
| CVSS3.0 : HIGH | |
| スコア | 8.1 |
|---|---|
| Vector | CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H |
| 攻撃元区分(AV) | ネットワーク |
| 攻撃条件の複雑さ(AC) | 高 |
| 攻撃に必要な特権レベル(PR) | 不要 |
| 利用者の関与(UI) | 不要 |
| 影響の想定範囲(S) | 変更なし |
| 機密性への影響(C) | 高 |
| 完全性への影響(I) | 高 |
| 可用性への影響(A) | 高 |
| CVSS2.0 : MEDIUM | |
| Score | 6.8 |
|---|---|
| Vector | AV:N/AC:M/Au:N/C:P/I:P/A:P |
| 攻撃元区分(AV) | ネットワーク |
| 攻撃条件の複雑さ(AC) | 中 |
| 攻撃前の認証要否(Au) | 不要 |
| 機密性への影響(C) | 低 |
| 完全性への影響(I) | 低 |
| 可用性への影響(A) | 低 |
| Get all privileges. | いいえ |
| Get user privileges | いいえ |
| Get other privileges | いいえ |
| User operation required | いいえ |
| Configuration1 | or higher | or less | more than | less than | |
| cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.1:*:*:*:*:*:*:* | |||||
| cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.5:*:*:*:*:*:*:* | |||||
| cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.6:*:*:*:*:*:*:* | |||||
| execution environment | |||||
| 1 | cpe:2.3:h:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:-:*:*:*:*:*:*:* | ||||