CVE-2019-25723
| Summary |
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
|
| Publication Date |
June 3, 2026, 5:16 a.m. |
| Registration Date |
June 4, 2026, 4:15 a.m. |
| Last Update |
June 3, 2026, 5:16 a.m. |
|
CVSS3.1 : MEDIUM
|
| スコア |
4.0
|
| Vector |
CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L |
| 攻撃元区分(AV) |
ネットワーク |
| 攻撃条件の複雑さ(AC) |
高 |
| 攻撃に必要な特権レベル(PR) |
不要 |
| 利用者の関与(UI) |
不要 |
| 影響の想定範囲(S) |
変更あり |
| 機密性への影響(C) |
なし |
| 完全性への影響(I) |
なし |
| 可用性への影響(A) |
低 |
Related information, measures and tools