NVD Vulnerability Detail
Search Exploit, PoC
CVE-2019-25724
Summary

Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.

Publication Date June 3, 2026, 5:16 a.m.
Registration Date June 4, 2026, 4:15 a.m.
Last Update June 3, 2026, 5:16 a.m.
CVSS3.1 : MEDIUM
スコア 6.5
Vector CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
攻撃元区分(AV) 隣接
攻撃条件の複雑さ(AC)
攻撃に必要な特権レベル(PR) 不要
利用者の関与(UI) 不要
影響の想定範囲(S) 変更なし
機密性への影響(C) なし
完全性への影響(I) なし
可用性への影響(A)