NVD Vulnerability Detail
Search Exploit, PoC
CVE-2019-6538
Summary

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.

Publication Date March 26, 2019, 7:29 a.m.
Registration Date Jan. 26, 2021, 11:43 a.m.
Last Update Nov. 21, 2024, 1:46 p.m.
CVSS3.1 : MEDIUM
スコア 6.5
Vector CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N
攻撃元区分(AV) 隣接
攻撃条件の複雑さ(AC)
攻撃に必要な特権レベル(PR) 不要
利用者の関与(UI) 不要
影響の想定範囲(S) 変更なし
機密性への影響(C) なし
完全性への影響(I)
可用性への影響(A) なし
CVSS2.0 : LOW
Score 3.3
Vector AV:A/AC:L/Au:N/C:N/I:P/A:N
攻撃元区分(AV) 隣接
攻撃条件の複雑さ(AC)
攻撃前の認証要否(Au) 不要
機密性への影響(C) なし
完全性への影響(I)
可用性への影響(A) なし
Get all privileges. いいえ
Get user privileges いいえ
Get other privileges いいえ
User operation required いいえ
Affected software configurations
Configuration1 or higher or less more than less than
cpe:2.3:o:medtronic:mycarelink_monitor_firmware:24950:*:*:*:*:*:*:*
cpe:2.3:o:medtronic:mycarelink_monitor_firmware:24952:*:*:*:*:*:*:*
execution environment
1 cpe:2.3:h:medtronic:mycarelink_monitor:-:*:*:*:*:*:*:*
Configuration2 or higher or less more than less than
cpe:2.3:o:medtronic:carelink_monitor_firmware:2490c:*:*:*:*:*:*:*
execution environment
1 cpe:2.3:h:medtronic:carelink_monitor:-:*:*:*:*:*:*:*
Configuration3 or higher or less more than less than
cpe:2.3:o:medtronic:carelink_2090_firmware:-:*:*:*:*:*:*:*
execution environment
1 cpe:2.3:h:medtronic:carelink_2090:-:*:*:*:*:*:*:*
Configuration4 or higher or less more than less than
cpe:2.3:o:medtronic:amplia_crt-d_firmware:-:*:*:*:*:*:*:*
execution environment
1 cpe:2.3:h:medtronic:amplia_crt-d:-:*:*:*:*:*:*:*
Configuration5 or higher or less more than less than
cpe:2.3:o:medtronic:claria_crt-d_firmware:-:*:*:*:*:*:*:*
execution environment
1 cpe:2.3:h:medtronic:claria_crt-d:-:*:*:*:*:*:*:*
Configuration6 or higher or less more than less than
cpe:2.3:o:medtronic:compia_crt-d_firmware:-:*:*:*:*:*:*:*
execution environment
1 cpe:2.3:h:medtronic:compia_crt-d:-:*:*:*:*:*:*:*
Configuration7 or higher or less more than less than
cpe:2.3:o:medtronic:concerto_crt-d_firmware:-:*:*:*:*:*:*:*
execution environment
1 cpe:2.3:h:medtronic:concerto_crt-d:-:*:*:*:*:*:*:*
Configuration8 or higher or less more than less than
cpe:2.3:o:medtronic:concerto_ii_crt-d_firmware:-:*:*:*:*:*:*:*
execution environment
1 cpe:2.3:h:medtronic:concerto_ii_crt-d:-:*:*:*:*:*:*:*
Configuration9 or higher or less more than less than
cpe:2.3:o:medtronic:consulta_crt-d_firmware:-:*:*:*:*:*:*:*
execution environment
1 cpe:2.3:h:medtronic:consulta_crt-d:-:*:*:*:*:*:*:*
Configuration10 or higher or less more than less than
cpe:2.3:o:medtronic:evera_icd_firmware:-:*:*:*:*:*:*:*
execution environment
1 cpe:2.3:h:medtronic:evera_icd:-:*:*:*:*:*:*:*
Configuration11 or higher or less more than less than
cpe:2.3:o:medtronic:maximo_ii_crt-d_and_lcd_firmware:-:*:*:*:*:*:*:*
execution environment
1 cpe:2.3:h:medtronic:maximo_ii_crt-d_and_lcd:-:*:*:*:*:*:*:*
Configuration12 or higher or less more than less than
cpe:2.3:o:medtronic:mirro_icd_firmware:-:*:*:*:*:*:*:*
execution environment
1 cpe:2.3:h:medtronic:mirro_icd:-:*:*:*:*:*:*:*
Configuration13 or higher or less more than less than
cpe:2.3:o:medtronic:nayamed_nd_icd_firmware:-:*:*:*:*:*:*:*
execution environment
1 cpe:2.3:h:medtronic:nayamed_nd_icd:-:*:*:*:*:*:*:*
Configuration14 or higher or less more than less than
cpe:2.3:o:medtronic:primo_icd_firmware:-:*:*:*:*:*:*:*
execution environment
1 cpe:2.3:h:medtronic:primo_icd:-:*:*:*:*:*:*:*
Configuration15 or higher or less more than less than
cpe:2.3:o:medtronic:protecta_icd_and_crt-d_firmware:-:*:*:*:*:*:*:*
execution environment
1 cpe:2.3:h:medtronic:protecta_icd_and_crt-d:-:*:*:*:*:*:*:*
Configuration16 or higher or less more than less than
cpe:2.3:o:medtronic:secura_icd_firmware:-:*:*:*:*:*:*:*
execution environment
1 cpe:2.3:h:medtronic:secura_icd:-:*:*:*:*:*:*:*
Configuration17 or higher or less more than less than
cpe:2.3:o:medtronic:virtuoso_icd_firmware:-:*:*:*:*:*:*:*
execution environment
1 cpe:2.3:h:medtronic:virtuoso_icd:-:*:*:*:*:*:*:*
Configuration18 or higher or less more than less than
cpe:2.3:o:medtronic:virtuoso_ii_icd_firmware:-:*:*:*:*:*:*:*
execution environment
1 cpe:2.3:h:medtronic:virtuoso_ii_icd:-:*:*:*:*:*:*:*
Configuration19 or higher or less more than less than
cpe:2.3:o:medtronic:visia_af_icd_firmware:-:*:*:*:*:*:*:*
execution environment
1 cpe:2.3:h:medtronic:visia_af_icd:-:*:*:*:*:*:*:*
Configuration20 or higher or less more than less than
cpe:2.3:o:medtronic:viva_crt-d_firmware:-:*:*:*:*:*:*:*
execution environment
1 cpe:2.3:h:medtronic:viva_crt-d:-:*:*:*:*:*:*:*
Related information, measures and tools
Common Vulnerabilities List

JVN Vulnerability Information
複数の Medtronic 製品におけるアクセス制御に関する脆弱性
Title 複数の Medtronic 製品におけるアクセス制御に関する脆弱性
Summary

複数の Medtronic 製品には、アクセス制御に関する脆弱性が存在します。

Possible impacts 情報を改ざんされる可能性があります。
Solution

ベンダ情報および参考情報を参照して適切な対策を実施してください。

Publication Date March 21, 2019, midnight
Registration Date May 9, 2019, 12:26 p.m.
Last Update May 9, 2019, 12:26 p.m.
Affected System
Medtronic
CareLink 2090 ファームウェア 
CareLink Monitor ファームウェア 2490C
Claria CRT-D ファームウェア 
Compia CRT-D ファームウェア 
Concerto CRT-D ファームウェア 
Concerto II CRT-D ファームウェア 
Consulta CRT-D ファームウェア 
Evera ICD ファームウェア 
Maximo II CRT-D and LCD ファームウェア 
MyCareLink Monitor ファームウェア 24950
MyCareLink Monitor ファームウェア 24952
CVE (情報セキュリティ 共通脆弱性識別子)
CWE (共通脆弱性タイプ一覧)
ベンダー情報
その他
Change Log
No Changed Details Date of change
1 [2019年05月09日]
  掲載
May 9, 2019, 12:26 p.m.